Papers
Latest Papers
"Enabling Autonomous Colonoscopy Intervention Using a Robotic Endoscope Platform" is now published in the IEEE Transactions on Biomedical Engineering!
Books, Book Chapters, and Magazine Articles
- Rentschler, M., “Mech-ing Endoscopies Safer by Using Robots You Can Swallow,” Futurum, December, 2020. (,Materials available for FREE download)
- Rentschler, M., Oleynikov, D., “Will In vivo Robotics Be the Future of Minimally Invasive Surgery?” Med Tech Business Review. 1(4): 70-73, 2007.
- Rentschler, M., “Undergraduates Benefit From Conducting Research,” Nebraska Blueprint. Spring, 2001.
Journal Publications (peer-reviewed original articles)
- Formosa, G.A., Prendergast, J.M., Humbert, J.S., Rentschler, M.E., “Nonlinear Dynamic Modeling of a Robotic Endoscopy Platform on Synthetic Tissue Substrates,” ASME Journal of Dynamic Systems, Measurement and Control. 143(1): 011005 (11 pages), 2021. (Abstract, )
- Zhang, Q., Prendergast, J.M., Formosa, G.A., Fulton, M.J., Rentschler, M.E., “Enabling Autonomous Colonoscopy Intervention Using a Robotic Endoscope Platform,” IEEE Transactions on Biomedical Engineering. (Abstract, ,)
- Prendergast, J.M., Formosa, G.A., Fulton, M.J., Heckman, C., Rentschler, M.E., “A Real-Time State Dependent Region Estimator for Autonomous Endoscope Navigation,” IEEE Transactions on Robotics. (Abstract, ,)
- Ly, K., Kellaris, N., McMorris, D., Johnson, B.K., Acome, E., Sundaram, V., Naris, M., Humbert, J.S., Rentschler, M.E., Keplinger, C., Correll, N., “Miniaturized Circuitry for Capacitive Self-sensing and Closed-loop Control of Soft Electrostatic Transducers,” Soft Robotics.(Abstract, ,)
- Bowen, L.K., Johannes, K., Zuetell, E., Calahan, K., Edmundowicz, S.A., Long, R., Rentschler, M.E., “Patterned Enteroscopy Balloon Design Factors Influence Tissue Anchoring,” Journal of the Mechanical Behavior of Biomedical Materials. 111: 103966, 2020. (Abstract, )
- Johnson, B.K., Sundaram, V., Naris, M., Acome, E., Ly, K., Correll, N., Keplinger, C.M., Humbert, J.S., Rentschler, M.E., “Identification and Control of a Nonlinear Soft Actuator and Sensor System,” IEEE Robotics and Automation Letters. 5(3): 3783-3790, 2020. (Abstract, ,)
- Formosa, G.A., Prendergast, J.M., Edmundowicz, S.A., Rentschler, M.E., “Novel Optimization-Based Design and Surgical Evaluation of a Treaded Robotic Capsule Colonoscope,” IEEE Transactions on Robotics. 36(2): 545-552, 2020.(Abstract, , )
- Qi, Y., Calahan, K.N., Rentschler, M.E., Long, R., “Friction Between a Plane Strain Circular Indenter and a Thick Poroelastic Substrate,”Mechanics of Materials. 142: 103303, 2020. (Abstract, )
- Lauff, C., Knight, D., Kotys-Schwartz, D., Rentschler, M.E., "The Role of Prototypes in Communication between Stakeholders," Design Studies. 66: 1-34, 2020.(Abstract, )
- Johannes, K.G., Calahan, K.N., Qi, Y., Long, R., Rentschler, M.E., “Three-Dimensional Microscale Imaging and Measurement of Soft Material Contact Interfaces Under Quasi-Static Normal Indentation and Shear,” Langmuir. 35: 10725-10733, 2019.(Abstract, , )
- Sun, X., Yu, L., Rentschler, M.E., Wu, H., Long, R., "Delamination of a Rigid Punch from an Elastic Substrate Under Normal and Shear Forces,"Journal of the Mechanics and Physics of Solids. 122: 141-160, 2019.(Abstract, )
- Formosa, G.A., Prendergast, J.M., Peng, J., Kirkpatrick, D., Rentschler, M.E., “A Modular Endoscopy Simulation Apparatus (MESA) for Robotic Medical Device Sensing and Control Validation,” IEEE Robotics and Automation Letters. 3(4): 4054-4061, 2018.(Abstract, , )
- Lauff, C., Weidler-Lewis, J., O’Connor, K., Kotys-Schwartz, D., Rentschler, M.E., "Prototypes as Intermediary Objects for Design Coordination in First-Year Design Courses," International Journal of Engineering Education. 34(3): 1085-1103, 2018. (Abstract, )
- Prendergast, J.M., Formosa, G.A., Rentschler, M.E., "A Platform for Developing Robotic Navigation Strategies in a Deformable, Dynamic Environment," IEEE Robotics and Automation Letters. 3(3): 2670-2677, 2018.(Abstract, , )
- Lauff, C., Kotys-Schwartz, D., Rentschler, M.E., "What is a prototype? What are the roles of prototypes in companies?" ASME Journal of Mechanical Design. 140(6): 061102 (12 pages), 2018.(Abstract, )
- Kern, M.D., Long, R. Rentschler, M.E., “A Representative Volume Element Model for the Adhesion between a Micro-Pillared Surface and a Compliant Substrate,” Mechanics of Materials. 119: 65-73, 2018.(Abstract, )
- Fankell, D.P., Regueiro, R.A., Kramer, E.A., Ferguson, V.L., Rentschler, M.E., "A Small Deformation Thermo-Poromechanics Finite Element Model and its Application to Arterial Tissue Fusion," ASME Journal of Biomechanical Engineering. 140(3): 031007 (11 pages), 2018.*Editor’s Choice Paper(Abstract, )
- Prendergast, J.M., Perry, A., Patel, V., Lindley, E., Rentschler, M.E., "Positioning Performance of Power and Manual Drivers in Posterior Spinal Fusion Procedures," Applied Bionics and Biomechanics. 2017: 7262841 (9 pages), 2017. (Abstract, , )
- Han, F., Yang, X., Deng, Y., Rentschler, M., Yang, D., Zhang, H., "SRAL: Shared Representative Appearance Learning for Long-Term Visual Place Recognition," IEEE Robotics and Automation Letters. 2(2): 1172-1179, 2017. (Abstract, )
- Kern, M., Qi, Y., Long, R., Rentschler, M.E., "Characterizing Adhesion between a Micro-Patterned Surface and a Soft Synthetic Tissue," Langmuir. 33(4): 854-864, 2017. (Abstract,, )
- Kramer, E.A., Cezo, J., Fankell, D. P., Taylor, K.D., Rentschler, M.E., Ferguson, V.L., "Strength and Persistence of Energy-Based Vessel Seals Rely on Tissue Water and Glycosaminoglycan Content," Annals of Biomedical Engineering. 44(11): 3421-3431, 2016. (Abstract,)
- Fankell, D., Kramer, E., Cezo, J., Ferguson, V.L., Taylor, K.D., Rentschler, M.E., "A Novel Parameter for Predicting Arterial Fusion and Cutting in Finite Element Models," Annals of Biomedical Engineering. 44(11): 3295-3306, 2016. (Abstract,)
- Sliker, L.J., Ciuti, G., Rentschler, M.E., Menciassi, A., "Frictional Resistance Model for Tissue-Capsule Endoscope Sliding Contact in the Gastrointestinal Tract,"Tribology International. 102: 472-484, 2016. (Abstract,)
- Kleck, C.J., Cullimore, I., LaFleur, M., Lindley, E., Rentschler, M.E., Burger, E.L., Cain, C.M.J., Patel, V.V., "A New 3-Dimensional Method for Measuring Precision in Surgical Navigation and Methods to Optimize Navigation Accuracy," European Spine Journal. 25(6): 1764-1774, 2016. (Abstract,)
- Francisco, M., Terry, B.S., Schoen, J.A., Rentschler, M.E., "An Intestinal Manometry Force Sensor for Robotic Capsule Endoscopy: An Acute, Multi-Patient In vivo Animal and Human Study,"IEEE Transactions on Biomedical Engineering. 63(5): 943-951, 2016. (Abstract,, )
- Anderson, N., Kramer, E., Cezo, J.D., Ferguson, V., Rentschler, M.E., "Bond Strength of Thermally Fused Vascular Tissue Varies with Apposition Force,"ASME Journal of Biomechanical Engineering. 137(12): 121010 (6 pages), 2015. (Abstract,)
- Sliker, L.J., Kern, M.D., Rentschler, M.E., "An Automated Traction Measurement Platform and Empirical Model for Evaluation of Rolling Micro-Patterned Wheels,"IEEE/ASME Transactions on Mechatronics. 20(4): 1854-1862, 2015. (Abstract,, )
- Cezo, J., Kramer, E., Schoen, J.A., Ferguson, V., Taylor, K., Rentschler, M.E., "Tissue Storage Ex Vivo Significantly Increases Vascular Fusion Bursting Pressure," Surgical Endoscopy. 29(7): 1999-2005, 2015. (Abstract,)
- Terry, B.S., Wang, X., Schoen, J.A., Rentschler, M.E., "A Preconditioning Protocol and Biaxial Mechanical Measurement of the Small Intestine,"International Journal of Experimental and Computational Biomechanics. 2(4): 293-309, 2014. (Abstract,)
- Kern, M., Ortega, J., Rentschler, M.E., "Soft Material Adhesion Characterization for In vivo Locomotion of Robotic Capsule Endoscopes: Experimental and Modeling Results," Journal of Mechanical Behavior of Biomedical Materials. 39: 257-269, 2014. (Abstract,)
- Zimkowski, M.M., Rentschler, M.E., Schoen, J.A., Mandava, N., Shandas, R., "Biocompatibility and Tissue Integration of a Novel Shape Memory Surgical Mesh for Ventral Hernia: In vivo Animal Studies,"Journal of Biomedical Materials Research, Part B Applied Biomaterials. 102(5): 1093-1100, 2014. ()
- Beccani, M., Di Natali, C., Sliker, L.J., Schoen, J., Rentschler, M.E., Valdastri, P. “Wireless Tissue Palpation for Intraoperative Detection of Lumps in Soft Tissue,” IEEE Transactions on Biomedical Engineering. 61(2): 353-361, 2014. *Cover Article (Abstract,)
- Cezo, J.D., Passernig, A., Ferguson, V., Taylor, K., Rentschler, M.E., "Evaluating Temperature and Duration in Arterial Tissue Fusion to Maximize Bond Strength,"Journal of Mechanical Behavior of Biomedical Materials. 30: 41-49, 2014. (Abstract,)
- Lyle, A.B., Terry, B.S., Schoen, J.A., Rentschler, M.E., "Preliminary Friction Force Measurements on Small Bowel Lumen when Eliminating Sled Edge Effects," Tribology Letters. 51(3): 377-383, 2013. (Abstract,)
- Cezo, J.D., Kramer, E., Taylor, K., Ferguson, V., Rentschler, M.E., "Temperature Measurement Methods during Direct Heat Arterial Tissue Fusion," IEEE Transactions on Biomedical Engineering. 60(9): 2552-2558, 2013. (Abstract,)
- Wang, X., Sliker, L.J., Qi, H., Rentschler, M.E., "A Quasi-static Model of Wheel-Tissue Interaction for Surgical Robotics,"Medical Engineering and Physics. 35(9): 1368-1376, 2013. (Abstract,)
- Zimkowski, M.M., Rentschler, M.E., Schoen, J.A., Rech, B.A., Mandava, N., Shandas, R., "Integrating a Novel Shape Memory Polymer into Surgical Meshes Decreases Placement Time in Laparoscopic Surgery: An In vitro and Acute In vivo Study,"Journal of Biomedical Materials Research Part A. 101(9): 2613-2620, 2013. ()
- Lyle, A.B., Luftig, J.T., Rentschler, M.E., "A Tribological Investigation of the Small Bowel Lumen Surface,"Tribology International. 62: 171-176, 2013. (Abstract,)
- Wang, X., Schoen, J.A., Rentschler, M.E., "A Quantitative Comparison of Soft Tissue Compressive Viscoelastic Model Accuracy,"Journal of Mechanical Behavior of Biomedical Materials. 20: 126-136, 2013. (Abstract,)
- Terry, B.S., Schoen, J.A., Rentschler, M.E., "Measurements of the Contact Force from Myenteric Contractions on a Solid Bolus,"Journal of Robotic Surgery. 7(1): 53-57, 2013. (Abstract,)
- Lindley, E., Zimkowski, M., Patel, V., Rentschler, M., "Pain Sensitivity Testing Using a Novel Computer-Controlled Pressure Algometer that Simultaneously Records Sympathetic Nervoous System Responses to Pain Stimuli,"Journal of Pain. 13(4): S9, 2012. (Abstract,)
- Terry, B.S., Passernig, A.C., Hill, M., Schoen, J.A., Rentschler, M.E., "Small Intestine Mucusal Adhesivity to In vivo Capsule Robot Materials,"Journal of Mechanical Behavior of Biomedical Materials. 15: 24-32, 2012. (Abstract,)
- Sliker, L.J., Rentschler, M.E., "The Design and Characterization of a Testing Platform for Quantitative Evaluation of Tread Performance on Multiple Biological Substrates,"IEEE Transactions on Biomedical Engineering. 59(9): 2524-2530, 2012. (Abstract,)
- Sliker, L., Kern, M., Schoen, J.A., Rentschler, M.E., "Surgical Evaluation of a Novel Tethered Robotic Capsule Endoscope using Micro-Patterned Treads,"Journal of Surgical Endoscopy. 26(10): 2862-2869, 2012. (Abstract,)
- Terry, B.S., Schoen, J.A., Rentschler, M.E., "Characterization and Experimental Results of a Novel Sensor for Measuring the Contact Force from Myenteric Contractions,"IEEE Transactions on Biomedical Engineering. 59(7): 1971-1977, 2012. (Abstract,)
- Terry, B.S., Mills, Z., Schoen, J.A., Rentschler, M.E., "Single Port Access Surgery with a Novel Magnet Camera System,"IEEE Transactions on Biomedical Engineering. 59(4): 1187-1193, 2012. (Abstract,)
- Rentschler, M.E., Hart, K.D., Mitchell, M.B., "Initial Design and Evaluation of a Pediatric Intra-Cardiac Camera System for Ventricular Septal Defects," ASME Journal of Medical Devices. 6(1): 011001-1-011001-9, 2012. (Abstract,)
- Terry, B.S., Schoen, J.A., Mills, Z., Rentschler, M.E., “Single Port Access Surgery with a Novel Port Camera System,” Surgical Innovation. 19(2): 123-129, 2012. (Abstract,)
- Terry, B.S., Lyle, A., Schoen, J.A., Rentschler, M.E., "Preliminary Mechanical Characterization of the Small Bowel for In vivo Robotic Mobility,"ASME Journal of Biomechanical Engineering. 133(9): 091010-1-091010-7, 2011. (Abstract,)
- Zimkowski, M., Lindley, E., Patel, V., Rentschler, M.E., "Design and Evaluation of a Computer-Controlled Pressure Algometer," ASME Journal of Medical Devices. 5(3): 031002-1-031002-6, 2011. (Abstract,)
- Sliker, L.J., Wang, X., Schoen, J.A., Rentschler, M.E., "Micropatterned Treads for In vivo Robotic Mobility,"ASME Journal of Medical Devices. 4(4): 041006-1-041006-8, 2010. (Abstract,, )
- Rentschler, M.E., Macdonald, S.A., “An Improved Brace Design for Boutonniere Deformity,” ASME Journal of Medical Devices. 4(2): 021002-1-021002-7, 2010. (Abstract,)
- Terry, B.S., Ruppert, A.D., Steinhaus, K.R., Schoen, J.A., Rentschler, M.E., “An Integrated Port Camera and Display System for Laparoscopy,” IEEE Transactions on Biomedical Engineering. 57(5): 1191-1197, 2010. (Abstract,)
- Platt, S.R., Hawks, J.A., Rentschler, M.E., “Vision and Task Assistance using Modular Wireless In Vivo Surgical Robots,” IEEE Transactions on Biomedical Engineering. 56(6): 1700-1710, 2009. ()
- Rentschler, M., Reid, J., "The Development of a Material Model and Wheel-Tissue Interaction for Simulating Wheeled Surgical Robot Mobility," Computer Methods in Biomechanics and Biomedical Engineering. 12(2): 239-248, 2009. ()
- Joseph, J.V., Oleynikov, D., Rentschler, M., Dumpert, J., Patel, H.R., “Microrobot Assisted Laparoscopic Urological Surgery in a Canine Model,” Journal of Urology. 180(5): 2202-2205, 2008. ()
- Rentschler, M., Baxter, B.T., "Screening Aortic Drug Treatments through Arterial Compliance Measurements,” Current Vascular Pharmacology. 6(4): 250-257, 2008. ()
- Soares, R.F., Allen, D.H., Kim, Y.R., Berthelot, C., Soares, J.B, Rentschler, M.E., “A Computational Model for Predicting the Effect of Tire Configuration on Asphaltic Pavement Life,” International Journal on Road Materials and Pavement Design. 9(2): 271-289, 2008. ()
- Lehman, A.C., Berg, K.A., Dumpert, J., Wood, N.A., Visty, A.Q., Rentschler, M.E., Platt, S.R., Farritor, S.M., Oleynikov, D., “Surgery with Cooperative Robots,” Computer Aided Surgery. 13(2): 95-105, 2008. ()
- Rentschler, M., Platt, S., Berg, K., Dumpert, J., Oleynikov, D., Farritor, S., “Miniature In vivo Robotics for Remote and Harsh Environments,” IEEE Transactions on Information Technology in Biomedicine. 12(1): 66-75, 2008. ()
- Rentschler, M., Baxter, B.T., “Medical Therapy Approach for Treating Abdominal Aortic Aneurysm,” Vascular. 15(6): 361-365, 2007. ()
- Rentschler, M., Iagnemma, K., Farritor, S., "Mechanical Design of Robotic In vivo Wheeled Mobility," ASME Journal of Mechanical Design. 129(10): 1037-1045, 2007. ()
- Rentschler, M., Dumpert, J., Platt, S., Farritor, S., Oleynikov, D., "Natural Orifice Surgery with an Endoluminal Mobile Robot,"Journal of Surgical Endoscopy. 21(7): 1212-1215, 2007. ()
- Rentschler, M., Dumpert, J., Platt, S., Iagnemma, K., Oleynikov, D., Farritor, S., "An In vivo Mobile Robot for Surgical Vision and Task Assistance," ASME Journal of Medical Devices. 1(1): 23-29, 2007. ()
- Rentschler, M., Baxter, B.T., "Pharmacological Approaches to Prevent Abdominal Aortic Aneurysm Enlargement and Rupture,"Annals of the New York Academy of Sciences. 1085: 39-46, 2006. ()
- Rentschler, M., Dumpert, J., Platt, S., Iagnemma, K., Oleynikov, D., Farritor, S., "Modeling, Analysis, and Experimental Study of In Vivo Wheeled Robotic Mobility,"IEEE Transactions on Robotics. 22(2): 308-321, 2006. ()
- Rentschler, M., Hover, F., Chryssostomidis, C., “System Identification of Open Loop Maneuvers Leads to Improved AUV Flight Performance,” IEEE Journal of Oceanic Engineering. 31(1): 200-208, 2006. ()
- Rentschler, M., Dumpert, J., Platt, S., Ahmed, S., Farritor, S., Oleynikov, D., “Mobile In Vivo Robots Provide Sole Visual Feedback for Abdominal Exploration and Cholecystectomy,” Journal of Surgical Endoscopy. 20(1): 135-138, 2006. ()
- Oleynikov, D., Rentschler, M., Hadzialic, A., Dumpert, J., Platt, S., Farritor, S., “Miniature Robots Can Assist in Laparoscopic Cholecystectomy,” Journal of Surgical Endoscopy. 19(4): 473-476, 2005. ()
Technical Briefs (peer-reviewed original articles)
- Prendergast, J.M., Perry, A., Rentschler, M.E., “Benchtop Testing of a Novel Robotic Capsule with Differential Drive Capabilities.” ASME Journal of Medical Devices.9(3): 030935(2 pp), 2015. *Selected as a Top Ten Paper(Abstract,, )
- Anderson, N., Kramer, E., Cezo, J., Ferguson, V., Rentschler, M.E., “Tissue Bond Strength as a Function of Applied Fusion Pressure,” ASME Journal of Medical Devices. 8(3): 030925(2 pp), 2014. ()
- Terry, B.S., Francisco, M., Schoen, J.A., Rentschler, M.E. “Sensor for Measuring the Contact Force from Human Myenteric Contractions for In vivo Robotic Capsule Endoscope Mobility,” ASME Journal of Medical Devices. 7(3): 030911(2 pp), 2013. ()
- Beccani, M., Di Natali, C., Rentschler, M.E., Valdastri, P. “Uniaxial Wireless Tissue Palpation Device for Minimally Invasive Surgery,” ASME Journal of Medical Devices. 7(2): 020919(3 pp), 2013. ()
Journal Publications (peer-reviewed review articles)
- Kramer, E.A., Rentschler, M.E., "Energy-Based Tissue Fusion for Sutureless Closure: Applications, Mechanisms, and a Potential for Functional Recovery," Annual Review of Biomedical Engineering. 20: 1-20, 2018. (Abstract, )
- Prendergast, J.M., Rentschler, M.E., "Towards Autonomous Motion Control in Minimally Invasive Robotic Surgery," Expert Review of Medical Devices. 13(8): 741-748, 2016. (Abstract,)
- Sliker, L.J., Ciuti, G., Rentschler, M.E., Menciassi, A., "Magnetically-driven Medical Devices: A Review,"Expert Review of Medical Devices. 12(6): 737-752, 2015. (Abstract,)
- Rentschler, M., Oleynikov, D., "Recent In vivo Surgical Robot and Mechanism Developments," Journal of Surgical Endoscopy. 21(9): 1477-1481, 2007. ()
- Lehman, A.C., Rentschler, M.E., Farritor, S.M., Oleynikov, D., “The Current State of Miniature In vivo Laparoscopic Robotics,” Journal of Robotic Surgery. 1(1): 45-49, 2007. ()
- Lehman, A.C., Rentschler, M.E., Farritor, S.M., Oleynikov, D., “Endoluminal Minirobots for Transgastric Peritoneoscopy,” Minimally Invasive Therapy and Allied Technologies. 15(6): 384-388, 2006. ()
- Rentschler, M., Platt, S., Dumpert, J., Farritor, S., Oleynikov, D., "In Vivo Laparoscopic Robotics,"International Journal of Surgery. 4(3): 167-171, 2006. ()
- Oleynikov, D., Rentschler, M.E., Dumpert, J., Platt, S.R., Farritor, S.M., “In Vivo Robotic Laparoscopy,” Surgical Innovation. 12(2): 177-181, 2005. ()
Conference Presentations with Publication (peer-reviewed original paper)
- Fulton, M.J., Prendergast, J.M., DiTommaso, E.R., Rentschler, M.E., “Comparing Visual Odometry Systems in Actively Deforming Simulated Colon Environments,” IEEE International Conference on Intelligent Robots and Systems (IROS), Las Vegas, NV, October, 2020. (Abstract, , )
- Formosa, G.A., Prendergast, J.M., Edmundowicz, S.A., Rentschler, M.E., “Novel Optimization-Based Design and Surgical Evaluation of a Treaded Robotic Capsule Colonoscope,” IEEE International Conference on Robotics and Automation, Paris, France, June, 2020.(Abstract, , ,)
- Johnson, B.K., Sundaram, V., Naris, M., Acome, E., Ly, K., Correll, N., Keplinger, C.M., Humbert, J.S., Rentschler, M.E., “Identification and Control of a Nonlinear Soft Actuator and Sensor System,” IEEE International Conference on Soft Robotics (RoboSoft), Yale University, April, 2020.(Abstract, , ,)
- Schunk, C., Pearson, L., Acome, E., Morrissey, T., Correll, N., Keplinger, C.M., Rentschler, M.E., Humbert, J.S., “System Identification and Closed-Loop Control of a Hydraulically Amplified Self-Healing Electrostatic (HASEL) Actuator,” IEEE International Conference on Intelligent Robots and Systems, Madrid, Spain, October, 2018.
- Prendergast, J.M., Formosa, G.A., Heckman, C.R., Rentschler, M.R., “Autonomous Localization, Navigation and Haustral Fold Detection for Robotic Endoscopy,” IEEE International Conference on Intelligent Robots and Systems, Madrid, Spain, October, 2018.(Abstract,,)
- Lauff, C.A., Kotys-Schwartz, D., Rentschler, M.E., “Prototyping Canvas: Building Purposeful Prototypes During Early Stage Design,” ASME International Design and Engineering Technical Conference, Quebec City, Canada, August, 2018.
- Lauff, C.A., Kotys-Schwartz, D., Rentschler, M.E., “Design Methods used during Early Stages of Product Development: Three Company Cases,” ASME International Design and Engineering Technical Conference, Quebec City, Canada, August, 2018.
- Lauff, C.A., Kotys-Schwartz, D., Rentschler, M.E., “What is a Prototype? Emergent Roles of Prototype from Empirical Work in Three Diverse Companies,” ASME International Design and Engineering Technical Conference, Cleveland, OH, August, 2017.
- Lauff, C.A., Kotys-Schwartz, D., Rentschler, M.E., “Perceptions of Prototypes: Pilot Study Comparing Students and Professionals,” ASME International Design and Engineering Technical Conference, Cleveland, OH, August, 2017.
- Han, F., Yang, X., Deng, Y., Rentschler, M., Yang, D., Zhang, H., "SRAL: Shared Representative Appearance Learning for Long-Term Visual Place Recognition," IEEE International Conference on Robotics and Automation (ICRA), Singapore, June, 2017.
- Ortega Alcaide, J., Pearson, L., Rentschler, M., "Design, Modeling and Control of a SMA-Actuated Biomimetic Robot with Novel Functional Skin," IEEE International Conference on Robotics and Automation (ICRA), Singapore, June, 2017. *Top 5 Paper by IEEE Spectrum Magazine (Abstract,,)
- Han, F., Yang, X., Deng, Y., Rentschler, M., Yang, D., Zhang, H., "Life-Long Place Recognition by Shared Representative Appearance Learning," Robotics: Science and Systems Conference, Ann Arbor, MI, June, 2016.
- O’Connor, K., Lauff, C., Kotys-Schwartz, D., Rentschler, M.E., “Learning and Identity at the Nexus of Practice: Mediated Discourse Analysis as a Methodology for Engineering Education Research,” ASEE Annual Conference and Exposition, Seattle, WA, June, 2015.
- Lauff, C., O’Connor, K., Kotys-Schwartz, D., Rentschler, M.E., “Comparing Organizational Structures: Two Case Studies of Engineering Companies,” ASEE Annual Conference and Exposition, Seattle, WA, June, 2015.
- Lauff, C., O’Connor, K., Kotys-Schwartz, D., Rentschler, M.E., “How is Design Organized? A Preliminary Study of Spatiotemporal Organization in Engineering Design,” IEEE Frontiers in Education Conference, Madrid, Spain, October, 2014.
- Lauff, C., Weidler-Lewis, J., O’Connor, K., Kotys-Schwartz, D., Rentschler, M.E., “Undergraduate to Professional Engineering Design: A Disconnected Trajectory?,” American Society for Engineering Education Zone IV Conference, Long Beach, CA, April, 2014.
- Wang. X., Di Natali, C., Beccani, M., Kern, M., Valdastri, P., Rentschler, M.E. “Novel Medical Wired Palpation Device: A Validation Study of Material Properties,” International Conference on Solid-State Sensors, Actuators and Microsystems – Transducers 2013 and Eurosensors XXVII, Barcelona, Spain, May, 2013.
- Beccani, M., Di Natali, C., Rentschler, M.E., Valdastri, P. “Wireless Tissue Palpation: Proof of Concept for a Single Degree of Freedom,” IEEE International Conference on Robotics and Automation, Karlsruhe, Germany, May, 2013.
- Sliker, L.J., Kern, M.D., Rentschler, M.E. “Preliminary Experimental Results and Modeling for a Four Degree of Freedom Automated Traction Measurement Platform for Quantitative Evaluation of In Vivo Robotic Capsule Colonoscopy Mobility Effectiveness,” IEEE International Conference on Robotics and Automation, Karlsruhe, Germany, May, 2013.
- Beccani, M., Di Natali, C., Rentschler, M.E., Valdastri, P. “Uniaxial Wireless Tissue Palpation Device for Minimally Invasive Surgery,” ASME Design of Medical Devices Conference, Minneapolis, MN, April, 2013.
- Terry, B.S., Francisco, M., Schoen, J.A., Rentschler, M.E. “Sensor for Measuring the Contact Force from Human Myenteric Contractions for In vivo Robotic Capsule Endoscope Mobility,” ASME Design of Medical Devices Conference, Minneapolis, MN, April, 2013.
- Terry, B.S., Schoen, J.A., Rentschler, M.E., "Preliminary Mechanical Characterization of the Small Bowel for In vivo Mobility,"ASME International Mechanical Engineering Congress and Exposition, Vancouver, British Columbia, Canada, November, 2010.
- Wang, X., Qi, H., Rentschler, M.E., "Analysis of Wheel-Tissue Interaction for In vivo Robotic Mobility,"ASME International Mechanical Engineering Congress and Exposition, Vancouver, British Columbia, Canada, November, 2010.
- Hart, K.D., Mitchell, M.B., Rentschler, M.E., "Initial Design and Evaluation of an Intra-Cardiac Camera System for Ventricular Septal Defects," ASME IDETC and CIE – Mechanisms and Robotics Conference, Montreal, Quebec, Canada, August, 2010.
- Sliker, L.J., Schoen, J.A., Rentschler, M.E., "Preliminary In vivo Capsule Crawler Mobility,"ASME IDETC and CIE – Mechanisms and Robotics Conference, Montreal, Quebec, Canada, August, 2010.
- Rentschler, M.E., Zable, J.A., “Beyond Capstone Design – Developing a New Graduate Design Pedagogy and Program,” Capstone Design Conference, «Ƶ, CO, June, 2010.
- Platt, S.R., Hawks, J.A., Rentschler, M.E., Redden, L., Oleynikov, D., Farritor, S., “Modular Wireless Wheeled In vivo Surgical Robots,” ASME DETC and CIE Conference - 30th Biennial Mechanisms and Robotics Conference, vol. 2, pp. 727-736, Brooklyn, NY, August, 2008.
- Rentschler, M., Dumpert, J., Platt, S., Oleynikov, D., Iagnemma, K., Farritor, S. “Mobile In Vivo Biopsy Robot,” IEEE International Conference on Robotics and Automation, pp. 4155-4160, Orlando, FL, May, 2006.
- Rentschler, M., Dumpert, J., Hadzialic, A., Platt, S., Oleynikov, D., Iagnemma, K., Farritor, S., “Theoretical and Experimental Analysis of In Vivo Wheeled Mobility,” Proceedings of the ASME DETC and CIE Conference - 28th Biennial Mechanisms and Robotics Conference, vol. 2, pp. 1241-1249, Salt Lake City, UT, October, 2004.
- Rentschler, M., Hover, F., Chryssostomidis, C., “Modeling and Control of an Odyssey III AUV through System Identification Tests”, 13th International Symposium on Unmanned Untethered Submersible Technology, Durham, NH, August, 2003.
- Farritor, S., Rentschler, M., “Robotic Highway Safety Markers”, American Society of Mechanical Engineers, Dynamic Systems and Control Division DSC, vol. 71, pp. 833-839, New Orleans, LA, November, 2002.
Conference Presentations with Publication (paper w/ peer-reviewed original abstract)
- Johannes, K., Bowen, L., Zuetell, E., Calahan, K., Long, R., Rentschler, M., “Bulk Experimental Data and Micro Scale Contact Modeling of Adhesion Mechanics of Highly Strained Silicone Micro-Patterned Surfaces in Contact With Soft Polyvinyl Chloride (PVC) Material,” ASME Verification and Validation Symposium, Baltimore, MD, May, 2020.
- Fankell, D., Regueiro, R., Rentschler, M.,“Simulating Arterial Tissue Fusion with a Large Deformation Thermo-Poromechanics Finite Element Model,” Engineering Mechanics Institute Conference, Cambridge, MA, May, 2018.
- Lauff, C., O’Connor, K., Kotys-Schwartz, D., Rentschler, M.E., “Engineering Design Process: More than the 'Standardized Loop',” American Society for Engineering Education Rocky Mountain Section Conference, Denver, CO, April, 2015.
- Kramer, E.A., Anderson, N.S. Taylor, K.D., Ferguson, V.L., Rentschler, M.E., "The Role of Glycosaminoglycans in Tissue Adhesion during Energy-Based Vessel Sealing," International Society for Optics and Photonics (SPIE) Photonics West, San Francisco, CA, February, 2015.
- Fankell, D., Kramer, E.A., Taylor, K.D., Ferguson, V.L., Rentschler, M.E., "A Novel Parameter for Predicting Arterial Fusion and Ablation in Finite Element Models,"International Society for Optics and Photonics (SPIE) Photonics West, San Francisco, CA, February, 2015.
- Cezo, J., Kramer, E., Taylor, K., Ferguson, V., Rentschler, M.E., "Tissue Fusion Bursting Pressure and the Role of Tissue Water Content," BiOSSPIE Photonics West, San Francisco, CA, February, 2013.
- Hawks, J., Rentschler, M., Farritor, S., Oleynikov, D., Platt, S., “A Modular Wireless In Vivo Surgical Robot with Multiple Surgical Applications,” Studies in Health Technology and Informatics - Medicine Meets Virtual Reality, vol. 142, pp. 117-121, IOS Press, Long Beach, CA, January, 2009.
- Hawks, J., Rentschler, M., Redden, L., Infanger, R., Dumpert, J., Farritor, S., Oleynikov, D., Platt, S., “Towards an In Vivo Wireless Mobile Robot for Surgical Assistance,” Studies in Health Technology and Informatics - Medicine Meets Virtual Reality, vol. 132, pp. 153-158, IOS Press, Long Beach, CA, January, 2008.
- Joseph, J.V., Boczko, J., Capello, S., Oleynikov, D., Dumpert, J., Rentschler, M., Patel, H.R., “Microrobots in Urology: Video Demonstration of Microrobot-Assisted Laparoscopic Prostatectomy and Nephrectomy in a Canine Model,” Journal of Endourology. 20: A311, 2006.
- Joseph, J.V., Oleynikov, D., Boczko, J., Rentschler, M., Dumpert, J., Patel, H.R., “Microrobots in urologic surgery: The next step in the evolution of robotic surgery,” Journal of Urology, 175(4): 344, 2006.
- Rentschler, M., Dumpert, J., Platt, S., Farritor, S., Oleynikov, D., “Mobile In Vivo Biopsy and Camera Robot,” Studies in Health Technology and Informatics - Medicine Meets Virtual Reality, vol. 119, pp. 449-454, IOS Press, Long Beach, CA, January, 2006.
- Rentschler, M., Dumpert, J., Platt, S., Farritor, S., Oleynikov, D., “Toward In Vivo Mobility,” Studies in Health Technology and Informatics - Medicine Meets Virtual Reality, vol. 111, pp. 397-403, IOS Press, Long Beach, CA, January, 2005.
- Rentschler, M., Oleynikov, D., Hadzialic, A., Dumpert, J., Platt, S., Farritor, S., “In Vivo Camera Robots Provide Improved Vision for Laparoscopic Surgery,” International Congress Series - Proceedings of Computer Assisted Radiology and Surgery Conference, vol. 1268, pp. 787-792, Chicago, IL, June, 2004.
- Rentschler, M., Hadzialic, A., Dumpert, J., Platt, S., Farritor, S., Oleynikov, D., “In Vivo Robots for Laparoscopic Surgery,” Studies in Health Technology and Informatics - Medicine Meets Virtual Reality, vol. 98, pp. 316-322, IOS Press, Newport Beach, CA, January, 2004.
Conference Presentations without Publication (peer-reviewed original abstract)
- Calahan, K., Qi, Y., Long, R., Rentschler, M.E., “Mapping Three-Dimensional Micromechanics for Investigation of Soft Tribology Mechanisms,” Gordon Research Conference – Science of Adhesion, Mount Holyoke College, MA, July, 2019.
- Calahan, K., Johannes, K., Long, R., Rentschler, M., “Mapping Three-dimensional Micro-mechanics between Micro-pillars and Soft Gel Substrates for Biomedical Application,” World Congress on Biomechanics, Dublin, Ireland, July, 2018.
- Johannes, K.G., Rentschler, M., “A Tunable Bio-Inspired Micro-Pillared Surface,” World Congress on Biomechanics, Dublin, Ireland, July, 2018.
- Rentschler, M., "Micro-Patterned Materials to Enable In vivo Robotic Capsule Endoscope Locomotion," MRS Spring Meeting, Invited Symposium Speaker on A Soft Future - From Electronic Skin to Robotics and Energy Harvesting, Phoenix, AZ, April, 2017. *Meeting Scientific Highlight
- Rentschler, M., "Towards Autonomous Robotic Capsule Endoscopy," IEEE International Conference on Intelligent Robots and Systems (IROS), Invited Symposium Speaker on Frontiers of Endoluminal Robotic Surgery, Daejeon, Korea, October, 2016.
- Kramer, E., Ferguson, V. L., Rentschler, M.E., "Biothermomechanics of Thermal Tissue Fusion: A Multi-Faceted Interaction,"World Congress on Biomechanics, Boston, MA, July, 2014.
- Cezo, J., Kramer, E., Schoen, J.A., Ferguson, V., Rentschler, M.E., "Tissue Storage Ex vivo Significantly Increases Vascular Fusion Bursting Pressure," The Society of American Gastrointestinal Endoscopic Surgeons, Salt Lake City, UT, April, 2014.
- Lindley, E.M., Spiegel, B., Zimkowski, M., Rentschler, M.E., Blount, T., Milligan, K., Burger, E.L., Patel, V.V., "A New Method For Clinically Assessing Pain," American Academy of Orthopaedic Surgeons Annual Meeting, New Orleans, LA, March, 2014.
- Kleck, C.J., LaFleur, M., Lindley, E.M., Clark, M., Burger, E.L., Cain, C.M.J., Rentschler, M.E., Razavi-Shearer, D., Patel, V.V., "Pedicle Screw Precision with 3-Dimensional Imaging, Navigation, and Measurement," SICOT Orthopaedic World Congress, Hyderabad, India, October, 2013.
- Lindley, E.M., Spiegel, B., Milligan, K., Zimkowski, M., Rentschler, M.E., Burger, E.L., Patel, V.V., "A New Method For Clinically Assessing Pain," SICOT Orthopaedic World Congress, Hyderabad, India, October, 2013.
- Lindley, E., Zimkowski, M., Patel, V., Rentschler, M.E., "Pain Sensitivity Testing Using A Novel Computer-Controlled Pressure Algometer That Simultaneously Records Sympathetic Nervous System Responses to Pain Stimuli," American Pain Society Annual Scientific Meeting, Honolulu, HI, May, 2012.
- Mills, Z., Terry, B., Schoen, J.A., Rentschler, M.E., "Single Port Access Surgery with a Novel Magnet Camera System,"ASME International Mechanical Engineering Congress and Exposition, Denver, CO, November, 2011 (IMECE2011-63963).
- Lyle, A., Terry, B., Schoen, J.A., Rentschler, M.E., "An Experimental Evaluation of Small Bowel Friction Forces when Eliminating Edge Effects,"ASME International Mechanical Engineering Congress and Exposition, Denver, CO, November, 2011 IMECE2011-63943).
- Kern, M., Sliker, L.J., Schoen, J.A., Rentschler, M.E., "Preliminary In vivo Capsule Endoscopic Mobility Using Micro-Patterned Treads,"ASME International Mechanical Engineering Congress and Exposition, Denver, CO, November, 2011 (IMECE2011-63723).
- Sliker, L.J., Schoen, J.A., Rentschler, M.E., "The Design and Experimental Evaluation of Micro-Patterned Treads for In vivo Robotic Mobility,"ASME International Mechanical Engineering Congress and Exposition, Denver, CO, November, 2011 (IMECE2011-63766).
- Cezo, J., Ferguson, V., Rentschler, M.E., "Computational Modeling of Direct Heat Tissue Fusion,"ASME International Mechanical Engineering Congress and Exposition, Denver, CO, November, 2011 (IMECE2011-62134).
- Terry, B.S., Schoen, J.A., Rentschler, M.E., "Development and Characterization of a Novel Sensor for Measuring the Radial Force from Myenteric Contractions,"ASME International Mechanical Engineering Congress and Exposition, Denver, CO, November, 2011 (IMECE2011-62121).
- Wang, X., Qi, H.J., Rentschler, M.E., "A Theoretical Study of Friction for In vivo Surgical Robotic Wheel-Tissue Interaction,"ASME International Mechanical Engineering Congress and Exposition, Denver, CO, November, 2011 (IMECE2011-62386).
- Wood, N., Lehman, A., Rentschler, M., Farritor, S., Oleynikov, D., “A Robotic Assistant for Surgical Dissection,” Medicine Meets Virtual Reality, Long Beach, CA, January, 2009.
- Lehman, A., Dumpert, J., Visty, A., Rentschler, M., Farritor, S., Oleynikov, D., "Towards Cooperative Miniature Robots for Natural Orifice Translumenal Endoscopic Surgery," Minimally Invasive Robotic Association (MIRA) Congress, Rome, Italy, January, 2008.
- Berg, K., Rentschler, M., Dumpert, J., Lehman, A., Platt, S., Oleynikov, D., Farritor, S., “Surgery with Cooperative Robots,” Medicine Meets Virtual Reality Conference, Long Beach, CA, February, 2007.
- Rentschler, M., Berg, K., Dumpert, J., Platt, S., Oleynikov, D., Farritor, S., “In vivo Robotics during the NEEMO 9 Mission,” Medicine Meets Virtual Reality Conference, Long Beach, CA, February, 2007.
- Rentschler, M., Dumpert, J., Lehman, A., Berg, K., Platt, S., Oleynikov, D., Farritor, S., “In Vivo Robots for Tele-Surgery During Long-Term Space Flight,” American Institute of Aeronautics and Astronautics Space Conference, San Jose, CA, September, 2006.
- Boczko, J., Capello, S., Oleynikov, D., Rentschler, M., Dumpert, J., Patel, H.R., Joseph, J.V., “Microrobots in Urology: Video Demonstration of Microrobot-Assisted Laparoscopic Prostatectomy and Nephrectomy in a Canine Model,” The 24th World Congress of Endourology, Cleveland, OH, August, 2006.
- Rentschler, M., Reid, J., “LS-DYNA Simulation of In vivo Surgical Robot Mobility,” The 9th International LS-DYNA Users Conference, Dearborn, MI, June, 2006.
- Rentschler, M., Dumpert, J., Platt, S., Farritor, S., Oleynikov, D., “Natural Orifice Surgery with an Endoluminal Mobile Robot,” The Society of American Gastrointestinal Endoscopic Surgeons, Dallas, TX, April 2006.
- Rentschler, M., Dumpert, J., Platt, S., Iagnemma, K., Oleynikov, D., Farritor, S., "In vivo Mobile Surgical Robotic Task Assistance,"Design of Medical Devices Conference, Minneapolis, MN, April, 2006.
- Rentschler, M., Dumpert, J., Platt, S., Farritor, S., Oleynikov, D., “Mobile In Vivo Robots Can Assist In Abdominal Exploration,” The Society of American Gastrointestinal Endoscopic Surgeons, Fort Lauderdale, FL, April, 2005.
- Dumpert, J., Rentschler, M., Platt, S., Farritor, S., Oleynikov, D., “Improving In Vivo Robot Vision Quality,”Medicine Meets Virtual Reality, Long Beach, CA, January, 2005.
- Narazaki, K., Oleynikov, D., DiMartino, A., Rentschler, M., Stergiou, N., “Objective Quantification of Proficiency in Robotic Laparoscopy with Bimanual Inanimate Tasks,”Medicine Meets Virtual Reality, Long Beach, CA, January, 2005.
- Oleynikov, D., Rentschler, M., Hadzialic, A., Dumpert, J., Platt, S., Farritor, S., “Miniature Robots Can Assist in Laparoscopic Cholecystectomy,” The Society of American Gastrointestinal Endoscopic Surgeons, Denver, CO, March, 2004.
Conference Posters (peer-reviewed original paper)
- Prendergast, J.M., Perry, A., Rentschler, M.E., “Benchtop Testing of a Novel Robotic Capsule with Differential Drive Capabilities.” ASME Design of Medical Devices Conference. Minneapolis, MN, April, 2015. *3-in-5 Presentation Award Winner
- Lauff, C., Weidler-Lewis, J., O’Connor, K., Kotys-Schwartz, D., Rentschler, M., “Cognitive Ethnographies of Heterogeneous Engineering Design,” International Conference of the Learning Sciences (ICLS), «Ƶ, CO, June, 2014.
- Anderson, N., Kramer, E., Cezo, J., Ferguson, V., Rentschler, M.E., “Tissue Bond Strength as a Function of Applied Fusion Pressure,” ASME Design of Medical Devices Conference, Minneapolis, MN, April, 2014.
Conference Posters (peer-reviewed original abstract)
- Calahan, K., Qi, Y., Rentschler, M., Long, R., “Mapping Three-Dimensional Micromechanics between Micro-Pillars and Soft Gel Substrates,” Society of Engineering Science Technical Meeting, St. Louis, MO, October, 2019.
- Johannes, K., Bowen, L., Calahan, K., Zuetell, E., Long, R., Rentschler, M.E., “Tunable Contact Properties of Highly Strained Hyperelastic Micro-Patterned Surfaces Contacting Soft Substrates,” Gordon Research Conference – Science of Adhesion, Mount Holyoke College, MA, July, 2019.
- Bowen, L., Johannes, K., Zuetell, E., Rentschler, M.E., “Identifying Friction Mechanisms of Deformable Microtextured Surfaces for Improved Anchoring during Balloon Endoscopy,” Gordon Research Conference – Science of Adhesion, Mount Holyoke College, MA, July, 2019.
- Anderson, N., Kramer, E., Cezo, J.D., Ferguson, V. L., Rentschler, M.E., "Tissue Bond Strength and Intraluminal Temperature as a Function of Applied Fusion Pressure,"World Congress on Biomechanics, Boston, MA, July, 2014.
- Kern, M., Ortega, J., Rentschler, M.E., "Soft Material Adhesion Characterization for In vivo Locomotion: Experimental and Modeling Results,"World Congress on Biomechanics, Boston, MA, July, 2014.
- Fankell, D., Ferguson, V.L., Taylor, K.D., Rentschler, M.E., "Impact of Heater Location and Ambient Conditions on Fusion Strength during Direct Heat Fusion,"World Congress on Biomechanics, Boston, MA, July, 2014.
- Cezo, J., Anderson, N., Kramer, E., Taylor, K., Rentschler, M., Ferguson, V., “Tissue Hydration Influences Bursting Pressure of Fused Arteries,” ASME Summer Bioengineering Conference, Sunriver, Oregon, June, 2013.
- Cezo, J., Ferguson, V., Taylor, K., Rentschler, M., “Measurement of Bond Strength of Direct Heat Tissue Fusion in Arteries,” ASME Summer Bioengineering Conference, Fajardo, Puerto Rico, June, 2012.
- Zimkowski, M.M., Lanning, C., Rech, B., Glang, R., Mandava, N., Mandava, N., Schoen, J., Rentschler, M., Shandas, R., “Mechanical Evaluation of Surgical Meshes Integrated with New Shape-Memory Polymer for Abdominal Wall Defect Repair,” American Hernia Society, March, 2012.
- Sliker, L., Kern, M.D., Schoen, J.A., Rentschler, M.E., “Biologically Inspired Micro-Patterned Treads for Robotic Capsule Endoscope Mobility,” Covidien R&D Summit, Westminster, CO, October, 2011.
- Terry, B.S., Schoen, J.A., Rentschler, M.E., “Experimental Measurements of the Radial Force from Myenteric Contractions on a Solid Bolus,” Covidien R&D Summit, Westminster, CO, October, 2011.
- Schoen, J.A., Mills, Z., Terry, B.S., Rentschler, M.E., “A SILS-Specific Camera System,” The Society of American Gastrointestinal Endoscopic Surgeons, San Antonio, TX, April, 2011.
- Schoen, J.A., Sliker, L., Rentschler, M.E., “A Locomotion System for Capsule Mobility,” The Society of American Gastrointestinal Endoscopic Surgeons, San Antonio, TX, April, 2011.
- Schoen, J.A., Terry, B.S., Mills, A., Rentschler, M.E., “Single Incision Laparoscopic Surgery with a Novel Port Camera System,” The Society of American Gastrointestinal Endoscopic Surgeons, Landover, MD, April, 2010.
- Macdonald, S., Rentschler, M., “An Improved Splint Design for Boutonniere Deformity,” ASME Summer Bioengineering Conference, Lake Tahoe, CA, June, 2009.
- Ruppert, A., Terry, B., Rentschler, M., “A Laparoscopic Camera-Enabled Cannula Port,” ASME Summer Bioengineering Conference, Lake Tahoe, CA, June, 2009.
- Lehman, A., Dumpert, J., Visty, A., Rentschler, M., Farritor, S., Oleynikov, D., "Toward Natural Orifice Surgery with Cooperative Miniature Robots," The Society of American Gastrointestinal and Endoscopic Surgeons, Philadelphia, PA, April, 2008.
- Rentschler, M., Baxter, B.T., “Preliminary Murine Aortic Tissue Material Properties from Pressure-Diameter Experiments,” ASME Summer Bioengineering Conference, Keystone, CO, June, 2007.
- Platt, S., Rentschler, M., Dumpert, J., Farritor, S., Oleynikov, D., “In Vivo Robotic Cameras Can Enhance Imaging Capability During Laparoscopic Surgery,” The Society of American Gastrointestinal Endoscopic Surgeons, Fort Lauderdale, FL, April, 2005.
Doctoral Dissertations
- Prendergast, J.M., Mobility, Navigation and Localization Towards Robotic Endoscopy, Doctoral Dissertation, University of Colorado, «Ƶ, CO, December, 2019.
- Lauff, C., Prototyping in the Wild: The Role of Prototypes in Companies, Doctoral Dissertation, University of Colorado, «Ƶ, CO, May, 2018.
- Fankell, D., A Thermo-Poromechanics Finite Element Model for Predicting Arterial Tissue Fusion, Doctoral Dissertation, University of Colorado, «Ƶ, CO, August, 2017.
- Kern, M., Adhesion Characterization of a Biologically Inspired Micro-Patterned Synthetic Material, Doctoral Dissertation, University of Colorado, «Ƶ, CO, December, 2016.
- Kramer, E., Thermal Fusion for Sutureless Closure: Devices, Composition, Methods, Doctoral Dissertation, University of Colorado, «Ƶ, CO, December, 2016.
- Sliker, L. Experimental Analysis and Modeling of Robot-Tissue Contact Mechanics for In Vivo Mobility, Doctoral Dissertation, University of Colorado, «Ƶ, CO, August, 2015.
- Cezo, J., Thermal Tissue Fusion of Arteries: Methods, Mechanisms, & Mechanics. Doctoral Dissertation, University of Colorado, «Ƶ, CO, August, 2013.
- Wang, X, Predictive Modeling of Tissue Mechanical Response for Rolling Motion, Doctoral Dissertation, University of Colorado, «Ƶ, CO, August, 2012.
- Terry, B., Mechanical Characterization of the Small Intestine for In vivo Robotic Capsule Endoscope Mobility, Doctoral Dissertation, University of Colorado, «Ƶ, CO, May, 2012.
- Rentschler, M., In Vivo Abdominal Surgical Robotics: Tissue Mechanics Modeling, Robotic Design, Experimentation, and Analysis, Doctoral Dissertation, University of Nebraska, Lincoln, NE, May, 2006.
Masters Theses
- Pearson, L., Modular Fabrication, Controller Development, and Sensor Development for a Hydraulically Amplified Self-Healing Electrostatic (HASEL) Actuator, Master’s Thesis, University of Colorado, «Ƶ, CO, December, 2018.
- Ortega, J., Design, Modeling and Control of a Soft SMA-actuated Biomimetic Robot with Functional Skin, Master's Thesis, University of Colorado, «Ƶ, CO, August, 2016.
- Kearns, J., A Comparison of Spring Stiffness Profiles For Use In A Passive-Elastic Leg Exoskeleton, Master's Thesis, University of Colorado, «Ƶ, CO, May, 2016.
- Ortega, J., Modeling Soft Material Adhesion for In vivo Robotic Locomotion, Master's Thesis – Basells Intern from Catalonia, Spain, February, 2014.
- Francisco, M., Migrating Motor Complex Force Sensor: Design, Fabrication and Clinical In vivo Measurements, Master's Thesis, University of Colorado, «Ƶ, CO, December, 2013.
- Hill, M., Design, Development, and Verification of an Automated Pain Measurement Device, Master's Thesis, University of Colorado, «Ƶ, CO, December, 2013.
- Lyle, A., Evaluation of Small Bowel Lumen Friction Forces for Applications in In vivo Robotic Capsule Endoscopy, Master's Thesis, University of Colorado, «Ƶ, CO, August, 2012.
- Hart, K., The Design and Evaluation of a Pediatric Cardiothoracic Camera System for Identification and Repair of Ventricular Septal Defects, Master's Thesis, University of Colorado, «Ƶ, CO, May, 2011.
- Mills, Z., Design and Experimental Evaluation of a Single Incision Laparoscopic Surgery Camera System for Minimally Invasive Surgery, Master's Thesis, University of Colorado, «Ƶ, CO, May, 2011.
- Zimkowski, M., Design and Preliminary Evaluation of a Computer-Controlled Pressure Algometer for Measuring Pain Sensitivity, Master's Thesis, University of Colorado, «Ƶ, CO, August, 2010.
- Ruppert, A., Design and Experimental Evaluation of an Integrated Cannula Visualization System for Laparoscopy, Master's Thesis, University of Colorado, «Ƶ, CO, August, 2009.
- Rentschler, M., Dynamic Simulation Modeling and Control of the Odyssey III Autonomous Underwater Vehicle, Master's Thesis, Massachusetts Institute of Technology, Cambridge, MA, June, 2003.
- Rentschler, M., Mobile Highway Construction Barrel Robots, Bachelor’s Thesis, University of Nebraska, Lincoln, NE, May, 2001.